MRI Safety Still a Concern, But National Guidance Lacking
Published on the July 27, 2012, DiagnosticImaging.com website
By Whitney L.J. Howell
Last October, the U.S. Food and Drug Administration (FDA) held a meeting to examine the biggest MRI safety concerns and outline best practices to reduce injuries associated with the scans. So far, however, the industry has received no national guidance on how to accomplish this goal.
But the lack of official instruction at the federal level doesn’t mean the push toward greater MRI safety — being recognized this week during MRI Safety Week — has stalled. Advancements are underway at the state level and in the accreditation process to prevent another tragedy, such as the 2001 Colombini case when a 6-year-old boy died when an oxygen tank flew across the room and struck him during the imaging process.
“Any protection that is implemented will benefit the 30 million Americans who get MRIs on an annual basis,” said Tobias Gilk, president and MRI safety director at Mednovus and senior vice president at the design and architecture firm Rad-Planning. “People are beginning to look harder at MRI risk factors as the power of our magnets continue to increase.”
State Regulations
Many states and the Joint Commission have implemented new regulations that strictly govern the design and construction of new MRI suites. These requirements do not mandate retrofitting existing installations.
“There are a growing number of states that are adopting this building code,” Gilk said. “Anyone who does MRI suite work — renovation, upgrade, equipment replacements, anything with a building permit — in virtually every jurisdiction must follow these state standards.”
To conform to the building regulations, health care engineers must meet eight standards, Gilk said. They must follow the American College of Radiology’s (ACR) 4-Zone Principles that addresse integrated access controls and screening practices. There must also be a clear line-of-sight between the operator’s console and patient inside the MRI machine, and designers must include demarcation lines to keep individuals with implantable medical devices far enough away. The presence of ferromagnetic-only detectors is required, and new MRI suites must also have exhaust fans and other protective measures to eliminate any cryogen that escapes into the imaging area.They must include an always-illuminated sign to remind staff that the MRI magnetic field is always active even without a patient.
These new design codes also require a clearly marked safe zone in which staff can use MR conditional equipment. Most importantly, Gilk said, designers and engineers must remember there is no cookie-cutter model for MRI suites. Each site is different and requires individual planning.
Accreditation
In a switch from years past, Gilk said, the Joint Commission is requiring far more documentation to prove you are making your MRI suite as safe as possible for patients. As part of the ACR’s 4-Zone Principles, the Commission now requires proof that you have established access control and provide proper staff supervision. You must also produce documentation that you have adequately trained your MRI personnel in safety procedures and protocols, he said.
“The Joint Commission wants to see evidence of screening protocols for implants and devices, clinical contraindications, and physical screenings,” Gilk said. “They want to see what you’re doing to make sure visitors aren’t carrying objects that will become dangers in the magnet room and what processes or tools are in place that accidents don’t occur.”
Requiring these measures — from an accreditation standpoint — is a giant step forward, he said.
To read the remainder of the story at its original location: http://www.diagnosticimaging.com/mri/content/article/113619/2092918
No comments yet.
-
Archives
- October 2019 (3)
- May 2018 (2)
- October 2017 (2)
- November 2016 (5)
- October 2016 (1)
- September 2016 (8)
- August 2016 (6)
- July 2016 (1)
- June 2016 (5)
- May 2016 (8)
- April 2016 (14)
- March 2016 (2)
-
Categories
-
RSS
Entries RSS
Comments RSS
Leave a Reply